P8.04 CAROTID INTIMA-MEDIA THICKNESS: COMPARISONS BETWEEN SEMI-AUTOMATIC EDGE DETECTION METHOD VERSUS A NEW AUTOMATIC ECHO TRACKING SYSTEM
- DOI
- 10.1016/j.artres.2010.10.086How to use a DOI?
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Rationale: The intima-media thickness (IMT) of the common carotid artery is a widely used clinical marker of cardiovascular diseases. To date, IMT is measured from longitudinal B-mode ultrasound images using manual or imaging analysis methods. Real-time automatic tracking of the lumen-intima and media-adventitia echoes using multi-echotracking technology could overcome some limits of manual and imaging analysis methods.
Material and methods: Common carotid intima-media thickness measured with a new multi-echotracking system (QIMT – Esaote - Italy) was compared to a semi-automatic edge detection imaging analysis method (IMT@televasc, www.televasc.fr, France) from B-mode ultrasound images.
Results: 93 scans were analyzed with both techniques on the same scans and from both carotids in 57 patients (age 55 +/− 17, 34 males). Mean QIMT was higher than IMT@televasc (690+/−173 vs 662+/−135, p<0.0001). Coefficient of variability for QIMT was 5% and 16 % for the IMT@televasc without significant inter-observer differences. Correlation coefficient were r= 0.657 (p<0.001, QIMT = IMT@televasc . 0.84 (+/−0.09) + 134+/−66 μm). Bland Altman plot did not showed systematic bias.
Discussion: QMIT gives more reproducible values of IMT than IMT@televasc with a significant correlation between the 2 methods. There was a systematic difference in absolute value between both techniques, values provided being higher with QIMT than with IMT@televasc. Conversion tables are needed for comparisons between echotracking technology and edge detection imaging analysis method for the routine use and follow up of the patients.
Acknowledgements
Mr Martin D, Esaote company and V de Germain, Televasc
Cite this article
TY - JOUR AU - G.P.N. Leftheriotis AU - R. Hamza AU - B. Levy AU - A. Taquet AU - L. Bressolette AU - M. Dauzat AU - P. Boutouyrie PY - 2010 DA - 2010/12/02 TI - P8.04 CAROTID INTIMA-MEDIA THICKNESS: COMPARISONS BETWEEN SEMI-AUTOMATIC EDGE DETECTION METHOD VERSUS A NEW AUTOMATIC ECHO TRACKING SYSTEM JO - Artery Research SP - 170 EP - 170 VL - 4 IS - 4 SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2010.10.086 DO - 10.1016/j.artres.2010.10.086 ID - Leftheriotis2010 ER -