3.2 A COMPARISON OF THE VICORDER APPARATUS WITH SPHYGMOCOR DEVICE FOR THE NON-INVASIVE ASSESSMENT OF AORTIC BLOOD PRESSURE: AN INVASIVE VALIDATION STUDY
- DOI
- 10.1016/j.artres.2010.10.167How to use a DOI?
- Open Access
- This is an open access article distributed under the CC BY-NC license.
Background: Aortic/central blood pressure (cBP) is an important determinant of cardiovascular risk. The SphygmoCor device applies a transfer function to radial BP waveforms calibrated to brachial systolic and diastolic BP (bSBP/bDBP) to estimate cBP parameters. Radial waveforms can be also calibrated to brachial mean BP (bMBP) and bDBP. Vicorder is a new cuff-based, operator-independent device which converts bBP waveforms to aortic waveforms to derive cBP.
Objective: to compare cBP estimated by non-invasive (Vicorder & SphygmoCor) devices to invasive cBP at cardiac catheterization.
Methods: Invasive BP (iBP) was measured in 33 patients (59±11 years, 63% males) undergoing diagnostic angiography, with a fluid-filled catheter at the aortic root. Simultaneous measurements were made using Vicorder and Sphygmocor. Brachial waveforms (Vicorder) were calibrated to oscillometric bSBP/bDBP; radial waveforms (Sphygmocor) were calibrated to oscillometric bSBP/bDBP and to bMBP/bDBP.
Results: Average (±SD) bSBP/bDBP was 145(±18)/81(±11) mmHg. iSBP/iDBP was 136/74(±18/9) mmHg; Vicorder-derived cSBP (137±17 mmHg) was in agreement with iSBP (mean ∆BP 0.3±8.0 mmHg, p=n.s.), while Vicorder-derived cDBP (81±11 mmHg) was higher than iDBP (mean ∆BP 7.0±7.4 mmHg, p<0.001). SphygmoCor bSBP/bDBP-calibrated cSBP (131±18 mmHg) under estimated iSBP (mean ∆BP −5.7±9.2 mmHg, p=0.002). SphygmoCor bMBP/bDBP-calibrated cSBP was higher than iSBP (141±17 mmHg, mean ∆BP 4.3±7.2, p=0.003).
Conclusions: Vicorder apparatus is highly accurate in non-invasive assessment of cBP compared with catheter-derived iSBP. SphygmoCor bSBP/bDBP-calibrated cSBP was lower than iSBP, while SphygmoCor bMBP/bDBP-calibrated cSBP was higher. Vicorder apparatus provides clinically useful values of cSBP.
Cite this article
TY - JOUR AU - G. Pucci AU - J. Cheriyan AU - A. Hubsch AU - S.S. Hickson AU - T. Watson AU - G. Schillaci AU - I.B. Wilkinson AU - C.M. McEniery PY - 2010 DA - 2010/12/02 TI - 3.2 A COMPARISON OF THE VICORDER APPARATUS WITH SPHYGMOCOR DEVICE FOR THE NON-INVASIVE ASSESSMENT OF AORTIC BLOOD PRESSURE: AN INVASIVE VALIDATION STUDY JO - Artery Research SP - 146 EP - 146 VL - 4 IS - 4 SN - 1876-4401 UR - https://doi.org/10.1016/j.artres.2010.10.167 DO - 10.1016/j.artres.2010.10.167 ID - Pucci2010 ER -